CertifiedBusiness Type: Trading Company CertifiedR&D Capacity: ODM Service Available & OEM Service Available CertifiedExport Year: 1 Year CertifiedNearest Port: Ningbo PortAverage Response Time: ≤9.30h CertifiedTerms of Payment: LC, T/T, PayPal, etc.
About Our Factory & Business Background
AddressN/a, Hong Kong, China
Plant Area20 square meters
Number of Employees2
Registered Capital10,000 HKD
International Commercial Terms(Incoterms)FOB, CIF, CFR, EXW
Average Lead TimePeak Season Lead Time: one month Off Season Lead Time: one month
Minimum Order Quantity1 Pc
Supply Chain Partners1
Our Production Capability & Technical Expertise
Main ProductsFreeze Dryer
Our Industry Experience & Global Business Record
Year of Establishment2025-05-06
Main MarketsNorth America, Europe
Repeat Buyers Choice30%~50%
Number of Foreign Trading Staff2
Overseas Agent/BranchNo
Our Certifications, Standards & Industry Recognition
Cooperated with Fortune 500No
Product Q&A
Q:How should I choose -60 C, -85 C or -105 C?
A:
-60 C is normally selected for routine aqueous samples. Choose -85 C for many solvent-containing or lower-eutectic samples, and -105 C for very low eutectic-point samples or dilute methanol/ethanol mixtures. The condenser should normally be 10-15 C colder than the sample eutectic or collapse temperature.
Q:Are the cabinet dimensions the same for all condenser temperatures?
A:
No. The -60 C version uses a more compact cabinet. The -85 C and -105 C configurations use a larger refrigeration cabinet. Request the configuration drawing before ordering.
Q:Can Pioneer 4L process peptide and protein samples?
A:
Yes. Pioneer is suitable for laboratory formulation screening and small-batch peptide, protein and biotechnology samples. The correct condenser temperature and cycle parameters depend on the formulation and critical product temperature.
Q:Which vacuum pump should I select?
A:
A rotary-vane pump is suitable for routine aqueous work. For solvents or acids, request a chemically resistant pump configuration, corrosion-resistant condenser option and appropriate secondary trap.
Send Inquiry
*To:SJ SCIENTIFIC LIMITED
*Content:
0/4000
Supplier replies will be sent to your registered email